Testo 330-2 LL - Flue Gas Analysers

With long-life gas sensors, H2-compensated CO-sensor, rech. battery and calibration protocol, graphic display, Flue gas probe (300mm x 8mm), carry case.
Manufacturer: Testo

Products

Picture of Testo 330-2 LL - Flue Gas Analyser (Pro set with Bluetooth)
Testo 330-2 LL - Flue Gas Analyser (Pro set with Bluetooth)
With long-life gas sensors, H2-compensated CO-sensor, rech. battery and calibration protocol, graphic display, Flue gas probe (300mm x 8mm), carry case, and Bluetooth function for use with testo combustion App
Part number: 0563 3372 72
£1,390.00 (ex VAT)
Picture of Testo 330-2 LL - Flue Gas Analyser (Pro set)
Testo 330-2 LL - Flue Gas Analyser (Pro set)
With long-life gas sensors, H2-compensated CO-sensor, rech. battery and calibration protocol, graphic display, Flue gas probe (300mm x 8mm), carry case
Part number: 0563 3372 70
£1,299.00 (ex VAT)
Picture of Testo 330-2 LL - Flue Gas Analyser (Pro set with printer)
Testo 330-2 LL - Flue Gas Analyser (Pro set with printer)
With long-life gas sensors, H2-compensated CO-sensor, rech. battery and calibration protocol, graphic display, Flue gas probe (300mm x 8mm), carry case, Testo printer, 6 x printer rolls
Part number: 0563 3372 71
£1,449.00 (ex VAT)

Description

The future of flue gas analysis has arrived with our comprehensive advanced analyser for the domestic and commercial Heating Engineer.

  • High-resolution colour display for the graphic representation of your measurement data
  • Extended measurement menus – such as gas pipe testing – for a comprehensive analysis of your heating system
  • Logger function for easy long-term recording of the measurement curve
  • Graphic processing of measurement data
  • Instrument diagnosis function uses easy „traffic light’’ presentation to enable comprehensive instrument diagnosis
  • Pressure measurement up to 300 mbar
  • User-defined fuels
  • Powerful memory management: 500,000 readings
  • USB interface for data readout to a PC software
  • ZIV (Central Guild Association) driver for all standard industry software packages
  • In CO measurement, from 8.000 ppm, automatic dilution is carried out up to min. 30,000 ppm CO
  • Integrated gas and draught zeroing without probe removal: The probe can remain in the flue during zeroing

Additional Bluetooth models available. Please call 01420 544 433 for further information.

For calibration of this instrument, please call us on 01420 544 433 to discuss


Memory

500.000 readings

Display

Colour graphic display with 240 x 320 Pixel

Storage temperature

-20 to +50 °C

Operating temperature

-5 to +45 °C

Weight

600 g

Dimensions

270 x 90 x 65 mm

Probe - Type K

Meas. Range

-40 to +1200 °C

Resolution

0.1 °C (-40 to 999,9 °C)

1 °C (remaining range)

Probe - CO

Meas. Range

0 to 500 ppm

Accuracy

±5 ppm (0 to 100 ppm)

±5% of mv (>100 ppm)

Resolution

1 ppm

Response time

Approx. 35 s

Probe - CO2

Meas. Range

0 to 1 Vol. %

0 to 10.000 ppm

Accuracy

±(50 ppm ±2% of mv) (0 to 5000 ppm)

Response time

Approx. 35 s

Probe - Differential pressure

Meas. Range

0 to 300 hPa

Accuracy

±0.5 hPa (0.0 to 50.0 hPa)

±1% of mv (50.1 to 100.0 hPa)

±1.5 % of mv (remaining range)

Resolution

0.1 hPa

Probe - Draught sensor

Meas. Range

-9.99 to +40 hPa

Accuracy

±0.02 hPa or ±5% of mv (-0.50 to +0.60 hPa)

±0.03 hPa (+0.61 to +3.00 hPa)

±1.5% of mv (+3.01 to +40.00 hPa)

Probe - Exhaust gas loss sensor

Meas. Range

0 to 99.9%

Resolution

0.1%

Probe - CO sensor

Meas. Range

0 to 8000 ppm

Accuracy

±10 ppm or ±10% of mv (0 to 200 ppm)

±20 ppm or ±5% of mv (201 to 2000 ppm)

±10% of mv (2001 to 8000 ppm)

Resolution

1 ppm

Probe - NO sensor (only applicable with NO upgrade sensor as optional extra)

Meas. Range

0 to 3000 ppm

Accuracy

±5 ppm (0 to 100 ppm)

±5% of mv (101 to 2000 ppm)

±10% of mv (2001 to 3000 ppm)

Resolution

1 ppm

Probe - O2 sensor

Meas. Range

0 to 21 Vol. %

Accuracy

±0.2 Vol. %

Resolution

0.1 Vol. %

Probe - Efficiency calculated

Meas. Range

0 to 120%

Resolution

0.1%

Probe - CO2 calculated

Meas. Range

0 to CO2 max

Accuracy

±0.2 Vol. %

Reaction time t90

< 40 s